This product consists of a PCD device and indicator strip. Mainly
use in testing sterilization quality of the sterilize package by medical
institutions.
Instruction:
1. Fold the indicator
strip then put it into the indicator cabin, close it and put the system into a
steel tray.
2. Put the tray in the
bottom of sterilizer to test.
3. Processing in
sterilization at 134℃
for 3.5mins or 121℃
for 15mins. If 4 color zones of the indicator strip is all turning to black
that means the sterilization is qualified. If one of the color zone is yellow
or brown that means the package is not regarded as asepsis. It needed to do the
B-D test again.
4. The tester decide whether
the sterilize package is qualified according to the result of the test.